Why are the beads from PTMScan® HS Kits aggregating against the sides of the tube?
Through in-house testing, we have found that the magnetic beads from our PTMScan HS kits aggregate and stick to the walls of some specific brands of tubes. Moreover, they cannot be washed off or spun off the sides of the tubes. This is an issue with specific brands of tubes themselves and not an inherent issue with the PTMScan bead product. Therefore, the following brands of polypropylene tubes are NOT recommended with magnetic bead PTMscan HS kits:
1. CellTreat #229441, polypropylene 1.5mL tube, non-sterile, non-pyrogenic
2. Denville Scientific (Harvard Bioscience) #C3171, 1.5mL Posi-Lock tubes, polypropylene
The following brand of microcentrifuge tubes is compatible with magnetic bead PTMscan HS kits:
1. USA scientific tubes #1615-5500
Last updated: 2024 年 3 月 14 日
Was this article helpful?
技术支持
电子邮件:[电子邮件受保护]
致电:4006-473287
联系授权经销商客户支持
电子邮件:[电子邮件受保护]
致电:4006-473287
传真:86-21-58356116
联系授权经销商联系销售
电子邮件:[电子邮件受保护]
致电:4006-473287
联系授权经销商